(regulation 1.4, definition of maximum permissible error )
In this Schedule:
"verification" includes reverification.
Part 2 Material measures
Table 1 Measures of Length
Item |
Lengths |
Maximum permissible error from zero to any scale mark for verification or in‑service inspection |
1 |
Not more than 500 mm |
±0.5 mm |
2 |
More than 500 mm but not more than 2 m |
±1 mm |
3 |
More than 2 m but not more than 100 m |
±0.05% |
Table 2 Non‑ferrous weights marked 'A'
Item |
Denomination |
Maximum permissible error (mg) |
||
|
|
Verification |
In‑service inspection |
|
|
|
|
Deficiency |
Excess |
1 |
1 mg |
+0.1 |
−0.05 |
+0.1 |
2 |
2 mg |
+0.2 |
−0.1 |
+0.2 |
3 |
5 mg |
+0.3 |
−0.15 |
+0.3 |
4 |
10 mg |
+0.4 |
−0.2 |
+0.4 |
5 |
20 mg |
+0.6 |
−0.3 |
+0.6 |
6 |
50 mg |
+0.9 |
−0.45 |
+0.9 |
7 |
100 mg |
+1.3 |
−0.65 |
+1.3 |
8 |
200 mg |
+2 |
−1 |
+2 |
9 |
500 mg |
+3 |
−1.5 |
+3 |
10 |
1 g |
+4 |
−2 |
+4 |
11 |
2 g |
+5.5 |
−2.75 |
+5.5 |
12 |
5 g |
+9 |
−4.5 |
+9 |
13 |
10 g |
+12.5 |
−6.25 |
+12.5 |
14 |
20 g |
+18 |
−9 |
+18 |
15 |
50 g |
+28 |
−14 |
+28 |
16 |
100 g |
+40 |
−20 |
+40 |
17 |
200 g |
+60 |
−30 |
+60 |
18 |
500 g |
+90 |
−45 |
+90 |
19 |
1 kg |
+130 |
−65 |
+130 |
20 |
2 kg |
+220 |
−110 |
+220 |
21 |
5 kg |
+280 |
−140 |
+280 |
22 |
10 kg |
+400 |
−200 |
+400 |
23 |
20 kg |
+560 |
−280 |
+560 |
Table 3 Non‑ferrous weights not marked 'A'
Item |
Denomination |
Maximum permissible error (mg) |
||
|
|
Verification |
In‑service inspection |
|
|
|
|
Deficiency |
Excess |
1 |
1 g |
+60 |
−30 |
+60 |
2 |
2 g |
+60 |
−30 |
+60 |
3 |
5 g |
+60 |
−30 |
+60 |
4 |
10 g |
+120 |
−60 |
+120 |
5 |
20 g |
+120 |
−60 |
+120 |
6 |
50 g |
+120 |
−60 |
+120 |
7 |
100 g |
+120 |
−60 |
+120 |
8 |
200 g |
+170 |
−85 |
+170 |
9 |
500 g |
+270 |
−135 |
+270 |
10 |
1 kg |
+380 |
−190 |
+380 |
11 |
2 kg |
+650 |
−325 |
+650 |
12 |
5 kg |
+850 |
−425 |
+850 |
13 |
10 kg |
+1 200 |
−600 |
+1 200 |
14 |
20 kg |
+1 700 |
−850 |
+1 700 |
Table 4 Iron Weights
Item |
Denomination |
Maximum permissible error (mg) |
||
|
|
Verification |
In‑service inspection |
|
|
|
|
Deficiency |
Excess |
1 |
100 g |
+240 |
−120 |
+240 |
2 |
200 g |
+340 |
−170 |
+340 |
3 |
500 g |
+540 |
−270 |
+540 |
4 |
1 kg |
+760 |
−380 |
+760 |
5 |
2 kg |
+1 300 |
−650 |
+1 300 |
6 |
5 kg |
+1 700 |
−850 |
+1 700 |
7 |
10 kg |
+2 400 |
−1 200 |
+2 400 |
8 |
20 kg |
+3 400 |
−1 700 |
+3 400 |
Table 5 Metric carat weights
Item |
Denomination (CM) |
Maximum permissible error (mg) |
||
|
Verification |
In‑service inspection |
||
|
|
Deficiency |
Excess |
|
0.005 |
+0.1 |
−0.05 |
+0.1 |
|
2 |
0.01 |
+0.1 |
−0.05 |
+0.1 |
3 |
0.02 |
+0.1 |
−0.05 |
+0.1 |
4 |
0.05 |
+0.1 |
−0.05 |
+0.1 |
5 |
0.1 |
+0.1 |
−0.05 |
+0.1 |
6 |
0.2 |
+0.15 |
−0.075 |
+0.15 |
7 |
0.5 |
+0.2 |
−0.1 |
+0.2 |
8 |
1 |
+0.2 |
−0.1 |
+0.2 |
9 |
2 |
+0.3 |
−0.15 |
+0.3 |
10 |
5 |
+0.5 |
−0.25 |
+0.5 |
11 |
10 |
+0.7 |
−0.35 |
+0.7 |
12 |
20 |
+1 |
−0.5 |
+1 |
13 |
50 |
+2 |
−1 |
+2 |
14 |
100 |
+2 |
−1 |
+2 |
15 |
200 |
+3 |
−1.5 |
+3 |
16 |
500 |
+5 |
−2.5 |
+5 |
Table 6 Conical measures
Item |
Capacity (L) |
Maximum permissible error (mL) |
|
|
|
Verification |
In‑service inspection |
0.5 |
+5 |
±5 |
|
2 |
1 |
+6 |
±6 |
3 |
2 |
+10 |
±10 |
4 |
4 |
+15 |
±15 |
5 |
5 |
+20 |
±20 |
6 |
10 |
+30 |
±30 |
7 |
20 |
+45 |
±45 |
Table 7 Cylindrical line measures
Item |
Capacity (L) |
Maximum permissible error (mL) |
||
|
Verification |
In‑service inspection |
||
|
|
Glass measures |
Metal measures |
|
1 |
0.5 |
±5 |
±5 |
±10 |
2 |
1 |
±10 |
±10 |
±20 |
3 |
2 |
±15 |
±15 |
±30 |
4 |
5 |
±30 |
±30 |
±60 |
5 |
10 |
±45 |
±45 |
±90 |
6 |
20 |
±70 |
±70 |
±140 |
Table 8 Cylindrical brim measures made of glass or metal for alcoholic liquor
Item |
Capacity (mL) |
Maximum permissible error (mL) |
|||
|
|
Verification |
In‑service inspection |
||
|
|
|
Glass measures |
Metal measures |
|
|
|
|
Deficiency |
Excess |
|
1 |
15 |
+1 |
0 |
+1 |
±1 |
2 |
30 |
+2 |
0 |
+2 |
±2 |
3 |
60 |
+3.5 |
0 |
+3.5 |
±3.5 |
Table 9 Cylindrical line measures made of metal for special purposes
Item |
Capacity (L) |
Purpose |
Maximum permissible error (mL) |
|
|
|
|
Verification |
In‑service inspection |
1 |
12 |
Ice cream |
±50 |
±100 |
2 |
45 |
Milk |
±150 |
±300 |
Table 10 Portable measures for potable liquids -- line measures
Item |
Type |
Maximum permissible error (mL) |
||
|
|
Verification |
In-service inspection |
|
|
|
|
Glass measures |
Metal measures |
1 |
Line measure |
±3% |
±3% |
±6 % |
Table 11 Portable measures for potable liquids -- brim measures
Item |
Type |
Maximum permissible error (mL) |
|||
|
|
Verification |
In‑service inspection |
||
|
|
|
Glass measures |
Metal measures |
|
|
|
|
Deficiency |
Excess |
|
1 |
Brim measure |
+6% |
0 |
+6% |
±6 % |
Division 4 Dispensing measures including pharmaceutical measures
Table 12 Conical dispensing measures
Item |
Scale mark (mL) |
Maximum permissible error at each scale mark (mL) for verification or in‑service inspection |
1 |
1 |
±0.08 |
2 |
2 |
±0.12 |
3 |
3 |
±0.16 |
4 |
4 |
±0.20 |
5 |
5 |
±0.25 |
6 |
6, 7, 8 |
±0.3 |
7 |
9 |
±0.4 |
8 |
10 |
±0.4 |
9 |
15 |
±0.5 |
10 |
20 |
±0.6 |
11 |
30 |
±0.8 |
12 |
40, 50 |
±1.0 |
13 |
60, 70, 80, 90 |
±1.5 |
14 |
100, 120, 140 |
±2.0 |
15 |
160, 180, 200 |
±3.0 |
Table 13 Beaker dispensing measures
Item |
Capacity (mL) |
Maximum permissible error at each scale mark on a particular measure (mL) for verification or in‑service inspection |
1 |
500 |
±5 |
2 |
1 000 |
±7 |
Table 14 Lubricating oil measures
Item |
Capacity (L) |
Maximum permissible error (mL) for verification or in‑service inspection |
1 |
0.5 |
+20 |
2 |
1 |
+30 |
Table 15 Graduated measuring cylinders
Item |
Capacity (mL) |
Maximum permissible error at each scale mark on a particular measure (mL) for verification or in‑service inspection |
1 |
5 |
±0.1 |
2 |
10 |
±0.2 |
3 |
25 |
±0.5 |
4 |
50 |
±1 |
5 |
100 |
±1 |
6 |
250 |
±2 |
7 |
500 |
±5 |
8 |
1 000 |
±10 |
9 |
2 000 |
±20 |
Division 1 Length‑measuring instruments
Table 1 Length‑measuring instruments
Item |
Type of indication |
Maximum permissible error |
|
|
Verification |
In‑service inspection |
|
1 |
Analog |
±0.5 % |
±1 % |
2 |
Digital |
±0.5 % + 0.5 scale interval |
±1 % + 0.5 scale interval |
Division 2 Area‑measuring instruments
Single‑measurement error -- instruments with analog indication
Table 2 Instruments with analog indication
Item |
Area of template (s) (dm 2 ) |
Maximum permissible error (dm 2 ) |
|
|
Verification |
In‑service inspection |
|
1 |
Not exceeding 25 |
±0.5 |
±1.0 |
2 |
Exceeding 25 |
±(0.5 + 1 dm2 for each additional 50 dm2 or part) |
±(1 + 2 dm 2 for each additional 50 dm 2 or part) |
Single‑measurement error -- instruments with digital indication
1 For instruments with digital indication add 0.5 scale interval to the maximum permissible error for an analog instrument.
Mean error
2 On analog and digital instruments the mean of 20 measurements must not differ from the denominated value of the template by more than half the maximum permissible error mentioned in table 2.
Note The test templates for measuring instruments with digital indication must have values that are an integral number of square decimetres.
The maximum permissible error for all scale marks on the dipsticks for verification and in-service inspection is ±1 scale interval.
Table 3 Farm milk tanks
Item |
Maximum capacity (L) |
Maximum scale interval (L) |
1 |
£ 1 000 |
2 |
2 |
1 000 £ 2 500 |
5 |
3 |
2 500 £ 5 000 |
10 |
4 |
5 000 £ 10 000 |
20 |
5 |
10 000 £ 25 000 |
50 |
Tanks used only for sullage
1 The maximum permissible error for verification or in-service inspection for each scale mark on a sight tube is ±0.5 scale interval.
Vehicle tanks except for sullage
2 The maximum permissible error applicable to a vehicle tank for verification or in-service inspection is:
( a ) for a tank with a capacity mark -- ±0.2% of the indicated volume; and
( b ) for a tank with a dipstick -- ±0.5 scale interval for each scale mark on the dipstick.
Division 5 Liquid‑measuring systems
Accuracy classes
1 Liquid‑measuring systems are classified into 5 accuracy classes as set out in table 4.
Table 4 Liquid‑measuring systems -- Accuracy classes
Item |
Accuracy class |
Field of application |
1 |
0.3 |
Measuring systems on pipeline |
2 |
0.5 |
All measuring systems if not differently stated elsewhere in this table, in particular: (a) fuel dispensers for motor vehicles (except LPG dispensers); and (b) measuring systems on road tankers for liquids of low viscosity; and (c) measuring systems for the unloading of ships' tanks and rail and road tankers; and (d) measuring systems for milk; and (e) measuring systems for loading ships; and (f) measuring systems for refuelling aircraft |
3 |
1.0 |
Measuring systems (except LPG dispensers) for liquefied gases under pressure measured at a temperature equal to or above −10° C LPG dispensers for motor vehicles Measuring systems normally in class 0.3 or 0.5 but used for liquids: (a) the temperature of which is less than −10° C or greater than 50°C; or (b) the dynamic viscosity of which is higher than 1 000 mPa.s; or (c) the maximum volumetric flow rate of which is not higher than 20 L/h |
4 |
1.5 |
Measuring systems for liquefied carbon dioxide Measuring systems (except LPG dispensers) for liquefied gases under pressure measured at a temperature below −10° C |
5 |
2.5 |
Measuring systems for liquids at a temperature below −153° C |
Maximum permissible errors
2 For volumes equal to or greater than 2 L, and subject to clauses 4 and 5, the maximum permissible error on volume indications are set out in table 5.
Table 5 Volumes equal to or greater than 2 L
Item |
Accuracy class |
Maximum permissible error |
||
|
Verification or in‑service inspection excluding conversion device |
A |
Conversion device |
|
1 |
0.3 |
± 0.2% |
± 0.3% |
± 0.1% |
2 |
0.5 |
± 0.3% |
± 0.5% |
± 0.2% |
3 |
1.0 |
± 0.6% |
± 1.0% |
± 0.4% |
4 |
1.5 |
± 1.0% |
± 1.5% |
± 0.5% |
5 |
2.5 |
± 1.5% |
± 2.5% |
± 1.0% |
3 For volumes less than 2 L, and subject to clauses 4 and 5, the maximum permissible errors, positive or negative, on volume indications are set out in table 6.
Table 6 Volumes less than 2 L
Item |
Measured quantity |
Maximum permissible errors |
1 |
³ 1 L < 2 L |
Value fixed in table 5 , applied to 2 L |
2 |
³ 0.4 L < 1 L |
2 x the value fixed in table 5 |
3 |
³ 0.2 L < 0.4 L |
2 x the value fixed in table 5 , applied to 0.4 L |
4 |
³ 0.1 L < 0.2 L |
4 x the value fixed in table 5 |
5 |
< 0.1 L |
4 x the value fixed in table 5 , applied to 0.1 L |
4 However, whatever the measured quantity may be, the magnitude of the maximum permissible error is the greater of the following 2 values:
(a) the absolute value of the maximum permissible error in table 5 or table 6;
(b) the minimum specified volume deviation under clause 5.
5 For minimum measured quantities equal to or greater than 2 L, the minimum specified volume deviation ( E min ) is calculated using the formula:
where:
"V min " is the minimum measured quantity.
"A" is the maximum permissible error specified in the column headed 'A' of table 5 for the accuracy class of the measuring system.
6 For minimum measured quantities less than 2 L, the minimum specified volume deviation is twice the value set out in table 6 and related to the column headed 'A' in table 5 .
Note The minimum specified volume deviation is an absolute maximum permissible error.
Accuracy of associated measuring instruments
7 Associated measuring instruments must exhibit an accuracy at least as good as the values set out in tables 7A, 7B and 7C for the accuracy class of the measuring system.
Table 7A Associated measuring instruments -- temperature
Item |
Accuracy classes of the measuring system |
Maximum permissible error on measuring temperature |
1 |
0.3 |
± 0.3°C |
2 |
0.5, 1.0 and 1.5 |
± 0.5°C |
3 |
2.5 |
± 1°C |
Table 7B Associated measuring instruments -- pressure
Item |
Accuracy classes of the measuring system |
Maximum permissible error on measuring pressure |
1 |
0.3, 0.5, 1.0, 1.5 and 2.5 |
Less than 1 MPa: ± 50 kPa Between 1 and 4 MPa: ± 5% More than 4 MPa: ± 200 kPa |
Table 7C Associated measuring instruments -- Density
Item |
Accuracy classes of the measuring system |
Maximum permissible error on measuring density |
1 |
0.3, 0.5, 1.0, 1.5 and 2.5 |
± 10 kg/m 3 |
Price computing devices
8 The price indicated must equal the price calculated for the volume and unit price indicated within the following maximum permissible errors as set out in table 8.
Table 8 Price computing devices
Item |
Scale interval |
Maximum permissible error for verification and in‑service inspection |
1 |
1 cent |
± 0.5 cent |
2 |
2 cents |
± 1.0 cent |
3 |
5 cents |
± 2.5 cents |
4 |
10 cents |
± 5.0 cents |
Beverage dispensers
Table 9 Beverage dispensers for alcoholic liquor
Item |
Capacity (mL) |
Maximum permissible error (mL) |
||
|
Verification |
In‑service inspection |
||
|
|
Deficiency |
Excess |
|
1 |
10 |
±0.5 mL |
−0.5 mL |
+1.0 mL |
2 |
15 |
±0.6 mL |
−0.6 mL |
+1.2 mL |
3 |
30 |
±1.0 mL |
−1.0 mL |
+2.0 mL |
4 |
60-100 |
±1.5 mL |
−1.5 mL |
+3.0 mL |
5 |
> 100 |
±1.5% of quantity dispensed |
−1.5% of quantity dispensed |
+3.0% of quantity dispensed |
Division 6 Weighing instruments
Class 1, 2, 3 or 4 non‑automatic instruments
1 The maximum permissible errors for increasing and decreasing loads, expressed in terms of verification scale interval ( e ), with an instrument adjusted to zero with ± 0.25 e at no load, are set out in table 10.
Table 10 Class 1, 2, 3 or 4 non‑automatic instruments
Item |
Load |
Maximum permissible error for verification and in-service inspection |
|||
Class 1 |
Class 2 |
Class 3 |
Class 4 |
||
1 |
0 < m £ 50 000 e |
0 < m £ 5 000 e |
0 < m £ 500 e |
0 < m £ 50 e |
±0.5 e |
2 |
> 50 000 e |
> 5 000 e |
> 500 e |
> 50 e, |
±1 e |
3 |
> 200 000 e |
> 20 000 e |
> 2 000 e |
> 200 e |
±1.5 e |
where e is the verification scale interval.
2 For digital indication or printing, a permissible error does not include the positive or negative error arising from rounding up or down to the nearest whole number of scale intervals.
Unclassified even‑arms scales
Table 11 Unclassified even‑arm scales
Item |
Capacity |
Maximum permissible error for verification and in-service inspection |
|||
|
|
Balances |
Beam scales |
Counter scales |
|
|
|
|
Class B |
Class C |
|
1 |
5 g |
±4 mg |
±10 mg |
‑ |
‑ |
2 |
25 g |
±6 mg |
±15 mg |
±60 mg |
‑ |
3 |
50 g |
‑ |
±20 mg |
‑ |
‑ |
4 |
100 g |
‑ |
±30 mg |
‑ |
‑ |
5 |
250 g |
‑ |
±60 mg |
±240 mg |
‑ |
6 |
500 g |
±12 mg |
±100 mg |
±400 mg |
±1.5 g |
7 |
1 kg |
‑ |
±150 mg |
±600 mg |
±2.5 g |
8 |
2 kg |
‑ |
±250 mg |
±1 g |
±3.5 g |
9 |
5 kg |
±70 mg |
±500 mg |
±2 g |
±6 g |
10 |
10 kg |
‑ |
± 1 g |
±4 g |
±8 g |
11 |
15 kg |
‑ |
±1.5 g |
±6 g |
±10 g |
12 |
25 kg |
±120 mg |
±2.5 g |
±10 g |
±15 g |
13 |
50 kg |
‑ |
±4.5 g |
±20 g |
±25 g |
3 The maximum permissible error for even‑arm scales must be half the amount specified in table 11 for loads not more than half capacity and the whole amount specified for loads more than half to maximum capacity.
Other unclassified instruments
4 The maximum permissible errors for self‑indicating weighing instruments and graduated non‑self‑indicating weighing instruments, with an instrument adjusted to zero within ±0.25 scale interval at no load must be:
(a) ±0.5 scale interval for the first 500 scale intervals; and
(b) ±1 scale interval for more than 500 but not more than 2 000 scale intervals; and
(c) ±1.5 scale intervals for more than 2 000 scale intervals.
Belt weighers
Table 12 Belt weighers
Item |
Class |
Maximum permissible error for or verification |
Maximum permissible error for in‑service inspection |
1 |
0.5 |
±0.25% |
±0.5% |
2 |
1.0 |
± 0.5% |
±1.0 % |
3 |
2.0 |
±1.0% |
±2.0% |
Catch weighers
5 The maximum permissible error for any load equal to or greater than the minimum capacity and equal to or less than the maximum capacity in automatic operation is set out in table 13.
Table 13 Automatic catch weighing instruments
Item |
Load ( m ) expressed in verification scale intervals (e ) |
Maximum permissible error for class Y (a) or Y (b) instruments -- verification or in‑service inspection |
|
|
Class Y (a) |
Class Y (b) |
|
1 |
0 < m £ 500 |
0 < m £ 50 |
± 1.5 e |
2 |
500 <m £ 2 000 |
50 <m £ 200 |
± 2 e |
3 |
2 000< m £ 10 000 |
200 < m £ 1 000 |
± 2.5 e |
Division 7 Automatic rail‑weighbridges
Accuracy classes
1 Automatic rail‑weighbridges are divided into the following 4 accuracy classes:
(a) 0.2;
(b) 0.5;
(c) 1;
(d) 2.
2 An instrument may be in an accuracy class for wagon weighing that is different from that for train weighing.
Maximum permissible errors -- weighing in motion
Table 14 Maximum permissible errors for weighing in motion
Item |
Accuracy class |
Percentage of weight of single wagon or total train, as appropriate |
|
|
|
Verification |
In‑service inspection |
1 |
0.2 |
± 0.10% |
± 0.2% |
2 |
0.5 |
± 0.25% |
± 0.5% |
3 |
1 |
± 0.50% |
± 1.0% |
4 |
2 |
± 1.00% |
± 2.0% |
Wagon weighing
3 The maximum permissible error for coupled or uncoupled wagon weighing is the greatest of the following values:
(a) the value calculated according to table 14, rounded to the nearest scale interval d ;
(b) the value calculated according to table 14, rounded to the nearest scale interval for the weight of a single wagon equal to 35% of the maximum wagon weight (as inscribed on the descriptive markings);
(c) 1 d .
Train weighing
4 The maximum permissible error for train weighing is the greatest of the following values:
(a) the value calculated according to table 14, rounded to the nearest scale interval d ;
(b) the value calculated according to table 14 for the weight of a single wagon equal to 35% of the maximum wagon weight (as inscribed on the descriptive markings) multiplied by the number of reference wagons in the train (not exceeding 10 wagons) and rounded to the nearest scale interval;
(c) 1 d for each wagon in the train, but not exceeding 10 d .
Maximum permissible errors -- static weighing
5 The maximum permissible errors on static weighing for increasing or decreasing loads must be the appropriate values set out in table 15.
Table 15 Static weighing
Item |
Maximum permissible error for verification and in-service inspection |
Load ( m ) expressed in numbers of scale intervals |
1 |
± 0.5 d |
0 £ m £ 500 |
2 |
± 1.0 d |
500 < m £ 2 000 |
3 |
± 1.5 d |
2 000 < m £ 10 000 |
Minimum capacity
6 The minimum capacity must not be less than 1 t, and not greater than the value of the result of the minimum wagon weight divided by the number of partial weighing.
Minimum wagon weight
7 The minimum wagon weight must not be less than 50 d .
Division 8 Grain protein measuring instruments
The maximum permissible errors at verification or in-service inspection are:
(a) wheat -- ± 0.4% at 11% moisture; and
(b) barley -- ± 0.5% at 0% moisture .
Division 9 Water Vending Machines
Table 16 Water vending machines at all volumes vended
Item |
Maximum permissible error |
||
|
Verification |
In‑service inspection |
|
|
|
Deficiency |
Excess |
1 |
± 1.5% |
− 1.5% |
+3% |
Division 10 Multi-dimensional measuring instruments
1 The maximum permissible error applicable to the measurement by the instrument of any of the 3 dimensions for verification and in-service inspection is +/- 1 d .
Table 1 Maximum permissible errors for water meters
Item |
Verification |
In‑service inspection |
1 |
±5% in the flow rate range where the flow rate ( Q) is greater than or equal to the minimum flow rate (Q 1 )and less than the transitional flow rate ( Q 2 ). ie: Q 1 £ Q < Q 2 ±2% in the flow rate range if the flow rate ( Q) is greater than or equal to the transitional flow rate ( Q 2 ) and less than or equal to the maximum flow rate ( Q 4 ). ie: Q 2 £ Q £ Q4 |
±4% in the flow rate range if the flow rate ( Q) is greater than or equal to 0.075 times the maximum continuous flow rate ( Q3) and less than or equal to the maximum flow rate (Q4). ie: 0.075 Q 3 £ Q £ Q 4 |
|
|
|